Biological Ventilation
Weighing / Sampling Booths
GMP FAT and SAT Certified
Booths designed for the fractionation, dispensing, and sampling of raw materials in GMP environments. Downward laminar flow with particle containment to simultaneously protect the operator and the product.
Quick facts
Available types
Recirculating / Non-recirculating / ATEX / Isolated
Downflow velocity
100 fpm
Capture grilles
Micro-perforated with 3 different capture zone velocities
Filtration
G4 Washable, M6 rigid bag, H14 long-life.
Complies with and is tested under
EN 1822
WHO REPORT 45
ISO 14644
Equipment anatomy
Designed component by component
MID MONITOR
Digital monitor with hot-wire sensor
DIFFUSER SCREENS
Screens to homogenize downflow velocities
TRIPLE FILTRATION
G4 Washable, M6 Rigid Bag, Long-life HEPA Filters.
CAPTURE GRILLES
304 Stainless Steel, micro-perforated with 3 capture velocity levels
UNIQUE SYSTEM THAT EXTENDS FILTER LIFESPAN
System that allows dust-laden air to shed particles before loading the filters, extending their useful life.
Regulatory traceability
We don't just say it complies. We prove it.
Standard
Requirements
How C4 complies
WHO INF 45
Must be a closed system that ensures containment and operator protection
Closed system with bidirectional flow, visual containment test that guarantees the containment of powders and reagents emitted in the work area. Flow uniformities, homogeneous capture throughout the equipment.
EN 1822
H14 Minimum efficiency of 99.995% for 0.3μm particles.
HEPA H14 filters, photometer tested in FAT and SAT.
WHO INF 45
Product protection
Exclusive capture zone for product protection, where external air is never directed to the product exposure area, visually demonstrated.
Engineering
Features that make the difference
The C4 Advantage
Regional manufacturing, not importation
40 years of manufacturing in LATAM allow us to adapt each piece of equipment to local regulations and provide support that an importer cannot match.
With C4
Minimum 10-year guarantee on spare parts availability
Warranty plans exceeding 3 years
Guaranteed operation under the conditions of each LATAM laboratory according to WHO INF 45
Guarantee of HEPA filters without replacement for at least 2 years
Imported equipment
No way to ensure spare parts locally for more than 2 years
Warranties limited to 1 year.
Performance guarantee of an older model
Filters requiring replacement every 6 months and difficult to source
Applications
Where this equipment works
PHARMA DISPENSING
PHARMA SAMPLING
INDUSTRIAL DISPENSING
INDUSTRIAL SAMPLING







How we work
From consultation to certified equipment
01
Diagnosis
02
Custom design
03
Manufacturing
04
Validation
Installation and certification at the factory and on-site.
Complementary solutions
You may also be interested in
Laminar flow cabinets
working with non-pathological processes
Process protection from the external environment.